What is a PSUR under the Medical Device Regulation (MDR) and Why Is It Important?
A PSUR is a document prepared by manufacturers of certain classes of medical devices that summarizes the results and conclusions drawn from the analysis of PMS data collected as part of the manufacturer's PMS plan. PMS is a crucial aspect of maintaining the safety and efficacy of medical devices once they are on the market and ensuring that real world data matches the expectations of the manufacturer based on bench testing and clinical studies. The PSUR includes:
- The conclusions of the benefit-risk analysis.
- Main findings from Post-Market Clinical Follow-up (PMCF).
- Data regarding the sales, usage, and population characteristics of the device.
Early dialogue with notified bodies, for example with our Structured Dialogue format, can be particularly valuable when defining the scope and methodology of PMCF activities. By clarifying expectations in advance, manufacturers can ensure that the data collected will meaningfully support the PSUR and facilitate a smoother review process.
What Is the Purpose of the PSUR?
The PSUR serves two primary purposes:
a. Identification and Evaluation of Changes in the Benefit-Risk Profile
The PMS is planned, implemented and reviewed by the manufacturer according to the requirements of the MDR, which takes a risk-based approach to define the frequency and depth of PMS. The analysis helps identify any changes in the benefit-risk profile of the device. The output of the analysis is particularly important for new devices where PMS gathers real world data that reflects the use of the device and where new information, or trends, emerges relating to the safety and performance of the device. Manufacturers must proactively collect PMS data and evaluate it to identify any safety or performance concerns. If adverse changes to the benefit-risk profile are detected, the manufacturer should evaluate this information in the context of the device's clinical evaluation and risk management processes. The PSUR ensures transparency by providing this critical information to the Notified Body and competent authorities involved in the device's conformity assessment.
b. Information on Corrective or Preventive Actions (CAPA)
The PSUR also serves as a tool for reporting any Corrective Actions or Preventive Actions (CAPA) related to the device, as required by Article 83(4) of the MDR. CAPAs in the PSUR include actions taken that:
• Are response to safety issues or performance concerns identified through PMS activities
• Directly impact product safety, performance, or quality.
Detailing the CAPAs in the PSUR ensures that any risks associated with the device are promptly addressed and communicated to the relevant authorities.
How Often Should You Update Your PSUR? PSUR Frequency Requirements by Device Class
- Class IIb and Class III Devices: Manufacturers are required to update their PSURs at least annually and promptly act upon any new data or trends. This means that within a year of the EU MDR certification date or the date of the 2023/607 transition letter, manufacturers must submit the PSUR to the Notified Body, even where the full MDR assessment has not been performed.
- Class IIa Devices: Manufacturers update the PSURs as necessary and at least every two years.
- Custom-Made Devices: For these unique devices, the PSUR forms part of the documentation required under Annex XIII of the MDR.
Note: PMS requirements of the EU MDR also apply to legacy devices under the MDD/AIMDD. For these devices, classification is under the Directive and the manufacturer’s issuance of the PSUR remains mandatory after the transition period until the end of the device lifetime defined in the technical documentation
PSUR Submission and Review Process under the Medical Device Regulation (MDR)
For Class III and implantable devices, manufacturers must submit PSURs electronically to the Notified Body involved in the conformity assessment. The Notified Body reviews the report and adds its evaluation. This information will be uploaded to the EUDAMED (European Database on Medical Devices), making it accessible to all competent authorities. For other classes of devices, manufacturers must make PSURs available to the Notified Body and, upon request, to competent authorities. The PSUR may be requested as part of routine surveillance; but is also likely to be requested by the EU competent authority or notified body where there are safety signals associated with a particular device or market information that suggests a new risk related to the type of device or its use.